Abstract
Background: Delta checks increase patient safety by identifying automated hematology analyzer errors. International standards and guidelines for the complete blood count (CBC) delta check method have not been established. We established an effective, practical CBC delta check method and criteria.

Methods: We assessed five delta check methods for nine CBC items (Hb, mean corpuscular volume, platelet count, white blood cell [WBC] count, and five-part WBC differential counts) using 219,804 blood samples from outpatients and inpatients collected over nine months. We adopted the best method and criteria and evaluated them using 42,652 CBC samples collected over two weeks with a new workflow algorithm for identifying test errors and corrections for Hb and platelet count.

Results: The median delta check time interval was 1 and 21 days for inpatients and outpatients (range, 1-20 and 1-222 days), respectively. We used delta values at 99.5% as delta check criteria; the criteria varied among the five methods and between outpatients and inpatients. The delta percent change (DPC)/reference range (RR) rate performed best as the delta check for CBC items. Using the new DPC/RR rate method, 1.7% of total test results exceeded the delta check criteria; the retesting and resampling rates were 0.5% and 0.001%, respectively.

Conclusions: We developed an effective, practical delta check method, including RRs and delta check time intervals, and delta check criteria for nine CBC items. The criteria differ between outpatients and inpatients. Using the new workflow algorithm, we can identify the causes of criterion exceedance and report correct test results.
All Author(s)
Min-Sun Kim ; Chan-Jeoung Park ; Seung Namgoong ; Seung-Il Kim ; Young-Uk Cho ; Seongsoo Jang
Issued Date
2023
Type
Article
Keyword
Automated hematology analyzerAutomationBlood cell countsDelta check criteriaDelta check methodQuality control
Publisher
대한진단검사의학회
ISSN
2234-3806 ; 2234-3814
Citation Title
Annals of laboratory medicine
Citation Volume
43
Citation Number
5
Citation Start Page
418
Citation End Page
424
Language(ISO)
eng
DOI
10.3343/alm.2023.43.5.418
URI
http://schca-ir.schmc.ac.kr/handle/2022.oak/1233
Appears in Collections:
진단검사의학과 > 1. Journal Papers
공개 및 라이선스
  • 공개 구분공개
파일 목록

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.