Abstract
The CT number corresponds to electron density and its influence on dose calculation was studied. Five kinds of CT scanners were used to obtain Images of electron density calibration phantom (Gammex RMI 467), Then the differences between CT numbers for each scanners were ${\pm}2\%$ In homogeneous medium and $9.5\%$ in high density medium. In order to Investigate the influence of CT number to dose calculation, patients' thoracic CT images were analyzed. The maximum dose difference was $0.48\%$ for each organ. It acquired the phantom Images inserted high density material in the water phantom. Comparing the doses calculated with CT Images from each CT scanner, the maximum dose difference was $2.1\%$ in 20 cm in depth. The exact density to CT number conversion according to CT scanner is required to minimize the uncertainty of dose depends on CT number Especially the each hospital with various CT scanners has to discriminate CT numbers for each CT scanner. Moreover a periodic quality assurance is required for reproducibility of CT number.
All Author(s)
Kwang Hwan Cho ; Suk Lee ; Sam Ju Cho ; Sangwook Lim ; Hyun Do Huh ; Chul Kee Min ; Byung-Chul Cho ; Yong Ho Kim ; Doo Ho Choi ; Eun Seog Kim ; Soo Il Kwon
Issued Date
2005
Type
Article
Keyword
Electron densityCT numberDose calculation
Publisher
한국의학물리학회
Korean Society of Medical Physics
ISSN
1226-5829 ; 2508-4445 ; 2508-4453
Citation Title
Korean Journal of Medical Physics
Citation Volume
16
Citation Number
4
Citation Start Page
161
Citation End Page
165
Language(ISO)
kor
URI
http://schca-ir.schmc.ac.kr/handle/2022.oak/1889
Appears in Collections:
방사선종양학과 > 1. Journal Papers
공개 및 라이선스
  • 공개 구분공개
파일 목록
  • 관련 파일이 존재하지 않습니다.

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.