SCHMC

Usefulness of newly modified fully covered metallic stent of 12mm in diameter and anti-migration feature for periampullary malignant biliary strictures: Comparison with conventional standard metal stent

Metadata Downloads
Abstract
BACKGROUNDS: Fully covered self-expandable metal stent (FCSEMS) are preferred for distal malignant biliary stricture (MBS). However, stent migration is a major adverse event of FCSEMS, especially for far distal MBS. We evaluated the usefulness of newly modified FCSEMS (M-FCSEMS) having 12mm in diameter and anti-migration feature to minimize stent migration compared with the conventional FCSEMS (C-FCSEMS). METHODS: Total 102 patients were enrolled between January 2015 and September 2017 in this prospective comparative study; 50 were allocated to the M-FCSEMS group and 52 to the C-FCSEMS group. The primary outcome was stent migration, and the secondary outcomes were other adverse events, stent occlusion rate, and stent patency during the follow-up period. RESULTS: The baseline characteristics of the two groups did not significantly differ. Endoscopic stent placement was technically successful in all patients. Stent migration occurred in 8.0% (4/50) of the patients in the M-FCSEMS group and 23.1% (12/52) of those in the C-FCSEMS group (P=0.036). The other adverse events, including stent-related pancreatitis and cholecystitis did not significantly differ between the two groups (P=0.415). Stent occlusion occurred in 23.9% (11/46) of the patients in the M-FCSEMS group and 37.5% (15/40) in the C-FCSEMS group (P=0.171). Stent patency was significantly longer in the M-FCSEMS group than in the C-FCSEMS group (228 vs 157days, P=0.048). CONCLUSIONS: Modified FCSEMS with 12-mm diameter and anti-migration feature significantly decreased the risk of stent migration and had longer patency compared with C-FCSEMS in patients with periampullary MBS.
All Author(s)
H. W. Lee ; J. H. Moon ; Y. N. Lee ; T. H. Lee ; M. H. Choi ; S. W. Cha ; Y. D. Cho ; S. H. Park
Issued Date
2019
Type
Article
Keyword
covered self-expandable metal stentmalignantbiliary strictureperiampullary cancerstentmigration
Publisher
Asian Pacific Association of Gastroenterology
ISSN
0815-9319 ; 1440-1746
Citation Title
Journal of Gastroenterology and Hepatology
Citation Volume
34
Citation Number
7
Citation Start Page
1208
Citation End Page
1213
Language(ISO)
eng
DOI
10.1111/jgh.14602
URI
http://schca-ir.schmc.ac.kr/handle/2022.oak/875
Appears in Collections:
소화기내과 > 1. Journal Papers
공개 및 라이선스
  • 공개 구분공개
  • 엠바고Forever
파일 목록
  • 관련 파일이 존재하지 않습니다.

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.